Assessing the feasibility of online SSVEP decoding in human walking using a consumer EEG headset

Abstract

Background

Bridging the gap between laboratory brain-computer interface (BCI) demonstrations and real-life applications has gained increasing attention nowadays in translational neuroscience. An urgent need is to explore the feasibility of using a low-cost, ease-of-use electroencephalogram (EEG) headset for monitoring individuals’ EEG signals in their natural head/body positions and movements. This study aimed to assess the feasibility of using a consumer-level EEG headset to realize an online steady-state visual-evoked potential (SSVEP)-based BCI during human walking.

Methods

This study adopted a 14-channel Emotiv EEG headset to implement a four-target online SSVEP decoding system, and included treadmill walking at the speeds of 0.45, 0.89, and 1.34 meters per second (m/s) to initiate the walking locomotion. Seventeen participants were instructed to perform the online BCI tasks while standing or walking on the treadmill. To maintain a constant viewing distance to the visual targets, participants held the hand-grip of the treadmill during the experiment. Along with online BCI performance, the concurrent SSVEP signals were recorded for offline assessment.

Results

Despite walking-related attenuation of SSVEPs, the online BCI obtained an information transfer rate (ITR) over 12 bits/min during slow walking (below 0.89 m/s).

Conclusions

SSVEP-based BCI systems are deployable to users in treadmill walking that mimics natural walking rather than in highly-controlled laboratory settings. This study considerably promotes the use of a consumer-level EEG headset towards the real-life BCI applications.

Background

Bridging the gap between laboratory brain-computer interface (BCI) demonstrations and real-life applications has gained increasing attention nowadays in translational neuroscience. A recent brain imaging study in mobile humans suggested that the brain switches to a different operating method while humans actively behave, move, walk, and orient in ecologically-valid environments [1]. That is, there might be significant differences in how the brain works in ecologically valid environments versus highly controlled laboratory environments [2, 3]. However, studies linking brain dynamics to cognitive functions have been majorly devoted to the stationary and tethered individuals, in which the participants underwent the experiments with highly-controlled settings in laboratories. They were usually instructed to avoid any task-irrelevant head/body movements. The laboratory-oriented research might have a generalizability issue when it is deployed in the real world. Therefore, studying the brain dynamics associated with naturalistic human behaviors is of great importance in translational neuroscience.

Promising advances in mobile electroencephalogram (EEG) system have largely progressed in naturalistic brain imaging. Wearing a lightweight mobile EEG headset allows EEG recording under naturalistic movements. Several previous studies have proved the efficacy of using either experimental prototype [46] or consumer-level EEG headsets [713], e.g. the MindWave headset (NeuroSky, Inc.) and the Emotiv EEG headset (Emotiv systems, Inc.) in cognitive neuroscience, neuroimaging, and BCI domains. Nevertheless, until recently only scattered studies [1418] employed such mobile brain sensing technology to explicitly assess EEG activities while participants freely and actively adapt to and interact with their naturalistic environments. Specifically, De Vos et al.[17] recently demonstrated the feasibility of a mobile auditory P300-based BCI system during natural walking. Nevertheless, the obtained BCI performance, i.e., information transfer rate (ITR), needed to be improved to satisfy performance requirement in practical applications.

The steady-state visual-evoked potential (SSVEP)-based BCI has become a popular communication channel that allows users to interact with environments/external devices due to its ease of use, minimal user training, large number of commands and high ITR [1921]. However, most of previous works were rooted in the SSVEP correlates of stationary and tethered subjects who were instructed to attentively gaze at the visual stimulation and avoid other gross movements. Recently, sparse works which adopted a laboratory-grade EEG device focused on an offline assessment of the SSVEP dynamics during walking [15, 18]. Alternatively to the use of lab-graded headgear, few studies noticed the need of realizing an SSVEP-based BCI using the consumer-level headset [1113]. Yet, these studies tested the efficacy with highly movement-controlled subjects. No study has reported the test of the consumer EEG headset for actuating the online SSVEP-based BCI with natural head/body positions and movements, i.e., in non-stationary, non-tethered subjects.

The main objective of this study was to assess whether or not a consumer-level headset can be used to perform online SSVEP decoding in human walking. To this end, the subjects were instructed to stand or walk on a speed-controllable treadmill (0.45, 0.89, and 1.34 meters per second (m/s)) in an attempt to initiate different levels of walking locomotion while using the online BCI. As compared to the standing condition, the methods and empirical results of moving subjects could shed the light on principles regarding the utility of a user-friendly and affordable commercial EEG headset for implementing real-life BCI applications towards naturalistic environments.

Methods

Participant

Seventeen healthy participants (14 males and 3 females; 22–32 years of age; mean age, 26.76 years) with normal or corrected-to-normal vision participated in this study. They were randomly selected from the student volunteers at University of California San Diego (UCSD). UCSD Human Research Protections Program approved this study. Each participant read and signed an informed consent before the experiment.

EEG recording

This study employed a 14-channel Emotiv EEG headset that sampled EEG signals at 128 Hz and band-pass filtered them between 0.2 and 45 Hz. The electrodes were positioned at AF3, F7, F3, FC5, T7, P7, O1, O2, P8, T8, FC6, F4, F8, and AF4 in accordance with the modified international 10–20 system. In addition, the angular velocities were concurrently recorded by the built-in two-channel gyroscope (Gyro-X: horizontal movement; Gyro-Y: vertical movement) in order to characterize the severity of the head movements during walking. The Emotiv Control Panel software provides visual monitoring of the electrode impedance. Once all electrodes had good contacts with the scalp, a laptop (Thinkpad X230, Lenovo Inc.) started to initiate the OpenViBE software [22] for connecting to the Emotiv headset, and then activate the BCILAB toolbox [23] to stream EEG signals for real-time frequency detection of the SSVEP signals.

Treadmill experiment setup

This study implemented a gaze-dependent SSVEP decoding system, which needed users to shift their visual focus to visual targets to elicit SSVEPs [21]. This study instructed participants to stand or walk on a treadmill at speeds of 0.45, 0.89, and 1.34 m/s (i.e., 1, 2, and 3 mile(s) per hour (MPH)) and intentionally gazed at a fixation cross or one of four repetitive black/white visual flickers (frequency: 9 Hz, 10 Hz, 11 Hz, and 12 Hz; size: 7 cm × 6 cm) concurrently presented on a 19-inch LCD monitor with a 60 Hz refresh rate (Figure 1). The paradigm can be used to implement a four-direction cursor control system. The generation of visual stimuli conformed to the approach that can be used to elicit SSVEPs at a flexible frequency [6]. The horizontal and vertical intervals between two stimuli were 19 cm and 17 cm respectively. The monitor was placed on a stand mounted above the control panel of the treadmill and adjusted to the same height as the participants’ eyes. The participants were instructed to hold the treadmill hand grip during standing and walking to maintain a constant viewing distance (~60 cm) between the subjects’ eyes and the monitor. Each subject participated in four sessions (each corresponded to one of four walking speeds from standing, 0.45 m/s, 0.89 m/s to 1.34 m/s) on the same day. Each session repeated a run 10 times. In each run, a verbal cue guided participants to shift their attention on the five targets (i.e., fixation cross and four flickers) sequentially. The participants were instructed to fixate at the fixation cross for three seconds and then shift their gaze to the flicker within 1.5 seconds after the verbal cue. The averaged time for detecting each target was 4.25 seconds (see below for more details on SSVEP detection). Each run lasted ~26 seconds in total. Thus, each session lasted around 4.3 minutes. Note the data within the periods of target-shifting were ignored in online SSVEP detection, and data recorded as participants gazed at the fixation cross were only used as baseline control in offline analysis. Each participant underwent a 4-session experiment (in the order of standing 0.45 m/s, 0.89 m/s to 1.34 m/s without randomization) with a minute(s) between-session rest to prevent visual and/or motor fatigue. During the rest, the signal qualities of electrodes were re-checked and adjusted via the Emotiv Control Panel. That is, without the need of removing the headset a syringe was used to drip a little more saline solution to the sponges of the electrodes with poor conductivity. The entire experiment for four sessions and between-session rests lasted less than 30 minutes.

Figure 1
figure 1

The illustration of experiment setup for SSVEP recordings.

Online SSVEP detection

The 14-channel EEG data in the 0.2-45 Hz frequency range (Emotiv default setting) were submitted to the process of frequency detection based on canonical correlation analysis (CCA) [19, 24] which has been proved highly efficient for enhancing the signal-to-noise ratio (SNR) of SSVEP signals [18, 19, 24]. CCA is a multivariate statistics method to maximize underlying correlation between two given multivariate signals x and y. To this end, CCA seeks the weight vectors w x and w y to project the input signals into X and Y, respectively, leading to the maximum canonical correlation ρ according to the following equation:

max W x , W y ρ X , Y = E w x T x y T w y E w x T x x T w x E w y T y y T w y

In this experiment, x refers to the n- second EEG signals (n is a self-regulated variable described below), and y refers to a set of reference signals generated as follows:

y= sin 2 πfn cos 2 πfn sin 2 π N h fn cos 2 π N h fn ,n= 1 f s , 2 f s ,, N f s

where f and fs are the target and sampling frequency, respectively. N h is the number of harmonics, and N is the number of sampling points. The CCA calculates the canonical correlation between the given EEG signals x and the reference signals y at each target frequency. The final decision of SSVEP frequency was made upon the maximum correlation among the reference signals. This study set the target frequency f as 9, 10, 11 and 12 Hz and the N h as two for generating template signals based on fundamental frequency and second harmonics with respect to the target frequency. Particularly, this study adopted the procedure of self-regulating data length [21] to sequentially accumulate the acquired EEG signals for improving the SNR of SSVEPs. This process used EEG data within several windows ranging from 2 to 8 seconds, with an increment of 0.25, for CCA calculation (e.g., 2, 2.25, 2.5… 8 s windows). Only if the detected frequency of the SSVEPs within four consecutive windows matched the target frequency, the detection pipeline would consider the target correctly classified. This detection pipeline was similar to the idea used in an adaptive P300 BCI [25]. Otherwise, the target would terminate as an incorrect decision. Note that, to eliminate the distraction from online feedback, this study did not present feedback to the subjects. The detection accuracy, decision time and ITR [26] were used to evaluate the SSVEP decoding performance. This study also saved the EEG data during the online experiments for further offline analysis.

Offline data analyses

The offline analyses aimed to explore the dynamics of the background EEG activity and the SSVEP amplitude acquired by the consumer headset under different walking speeds. The background EEG spectrum in the fixation-cross condition, which can be regarded as baseline power for the SSVEP conditions under each walking speed, was necessary to reveal the underlying brain activity associated with the walking locomotion. In addition, since this study adopted treadmill walking to systematically test the effects of different degrees of head movements on the SSVEPs, it is important to report the intensity and frequency of the head movements. To this end, we applied the spectrum analysis to the 2-channel gyroscope signals (horizontal and vertical axes) along different walking speeds. The detailed steps are described as follows. This study applied a 1 Hz high-pass finite impulse response (FIR) filter to the EEG signals to remove low-frequency drifts, resulting in a signal bandwidth of 1 to 45 Hz for the spectral analysis. The 128-point short-time Fourier transform (STFT) with a Hamming window of length 128 samples and 25% overlap was then applied to estimate the EEG spectrogram at a 1 Hz frequency resolution. The grand average power spectral density (PSD) of each channel can be derived for each condition. Note that, to dissociate the reactive SSVEP from the background EEG activities and walking related noises, this study estimated the relative power of SSVEPs by subtracting the power spectrum in the fixation-cross condition from those of the SSVEP conditions at each of the walking speeds. In addition, this study calculated the scalp distributions of the spectral power to form topographic maps using the EEGLAB toolbox [27]. The spectrogram of the recorded 2-channel Gyro signals was estimated using the same procedure as for EEG data (but without filtering). Lastly, this study employed a paired t-test to access the differences of SSVEP amplitudes and online performance between two walking conditions (standing, 0.45 m/s, 0.89 m/s, and 1.34 m/s).

Results

The impacts of walking movements on the EEG spectra

Figure 2 shows the averaged PSD of O1 and O2 channels across all subjects. In general, walking locomotion induced wideband noise to the EEG spectra, very likely arising from electrode shift with respect to the scalp and electromyography (EMG) activities from facial, neck, and scalp muscles during walking. This phenomenon led to power lifting proportional to walking speeds. Fast walking especially induced large augmentation of low-frequency power around 2 Hz. In addition, the SSVEP amplitudes under the standing and walking conditions exhibited salient peaks at the fundamental and second harmonics of the flickering frequencies, which appeared to monotonically attenuate as walking speed increased. To estimate SSVEP amplitudes under different walking speeds, this study alleviated the wideband noise by subtracting the power in the fixation-cross condition from that of the SSVEPs in the flickering conditions. Figure 3 shows that the SSVEP signals at the O1-O2 pair declined evidently as walking speed increased, leading to significant amplitude drops (p < 0.05) between standing versus walking (0.45, 0.89, and 1.34 m/s). No significant differences were found between the three walking speeds.Figure 4 illustrates an example of 2-channel gyroscope time course of a representative subject and averaged spectral profiles of 17 subjects across all conditions along the horizontal (Gyro-X) and the vertical (Gyro-Y) planes associated with walking locomotion under different walking speeds. As can be seen, compared with the standing condition, the low-frequency power (1 to 5 Hz) of the angular-velocity deviations tended to progressively augment as walking speed increased. Especially, the vertical head displacements (Gyro-Y) showed a big jump at 2 Hz while walking at 1.34 m/s. This result could be attributed to the fact that natural walking posture was typically accompanied by head bobbing (Gyro-Y) and head swaying (Gyro-X) while fast walking exhibited bigger head bobbing than head swaying.Figure 5 plots the topographic maps of the relative low-frequency power (1–5 Hz) and the SSVEP amplitudes (9–12 Hz) (derived from flickering minus fixation-cross) under different walking speeds. As aforementioned, the head movements during walking resulted in power increase at frequencies from 1 to 5 Hz. The topographic maps showed that the EEG power augmentations, which mainly distributed over the fronto-central areas versus the posterior areas, more and less positively correlated with the waking speed. Furthermore, the SSVEP maps clearly showed that the induced occipital SSVEP amplitudes tended to be weaker during fast walking. It is worth noting that since the topographic spectrum maps were interpolated on a sparse 14-channel montage with the absence of centro-parietal electrodes, the results did not lead to the implication of spectral changes in the sensorimotor cortex during walking.

Figure 2
figure 2

The average power spectral density of O1 and O2 channels across 17 subjects in response to the fixation-cross and visual targets flickering at 9, 10, 11, and 12 Hz while standing still or walking on the treadmill with the speeds of 0.45, 0.89, and 1.34 m/s.

Figure 3
figure 3

The averaged power of SSVEP signals at O1 and O2 channels across 17 subjects under different walking speeds after alleviating the wideband noise by subtracting the power in the fixation-cross condition from that of the SSVEPs in the flickering conditions. The thin lines indicate the significant difference between speeds (p < 0.05 or < 0.01), whereas the error bars represent the standard error of the SSVEP amplitudes.

Figure 4
figure 4

The left panel illustrates an example of 5-second time course of 2-channel gyroscope signals (GyroX: horizontal plane, GyroY: vertical plane) associated with different walking speeds from one representative subject. The right panel shows the averaged spectral profiles of 17 subjects across all stimulus conditions under different speeds.

Figure 5
figure 5

The topographical mapping of the relative low-frequency power (1–5 Hz) and SSVEP amplitude (9–12 Hz) (derived from flickering minus fixation-cross) under different walking speeds. The topographic legend indicates the channel locations of the 14-channel Emotiv EEG headset. The units of color bars represent the original EEG power in the dB scale. Note that these 14-channel interpolated maps (with the absence of the centro-parietal electrodes) do not imply spectral changes in the centro-parietal regions during walking.

The online SSVEP detection performance

Figure 6 shows the online SSVEP decoding performance in terms of detection accuracy (%), decision time (sec), and ITR (bits/min) at different walking speeds. In general, the classification accuracy was significantly higher than the chance level (25%) under all conditions. The averaged BCI performance progressively decayed as walking speed increased from 0.45 to 1.34 m/s (accuracy: 73.49 ± 26.52%, 71.83 ± 23.17%, 56.57 ± 16.56%; ITR: 14.04 ± 10.10, 12.37 ± 9.19, 5.53 ± 4.57; with the representation of the mean value plus/minus standard deviation), as compared to the standing condition (accuracy: 76.60 ± 21.74%, ITR: 14.38 ± 9.04). The significant drops (p < 0.05) in accuracy and ITR were only found between the 1.34 m/s condition versus the other three conditions. The decision time was found to slightly increase as walking speed increased, but there was no significant difference (p > 0.05) between speeds (standing: 4.34 ± 0.08 s; 0.45 m/s: 4.37 ± 0.10s; 0.89 m/s: 4.36 ± 0.09 s; 1.34 m/s: 4.38 ± 0.07 s).

Figure 6
figure 6

The online SSVEP decoding results in terms of averaged accuracy (%), decision time (sec), and ITR (bits/min) at the different walking speeds. The error bars represent the standard deviation of the online results.

Discussion

This study aimed to assess whether or not a consumer-level headset can be used to perform online SSVEP decoding for users in naturalistic positions, postures and movements. To this end, this study preliminarily tested the validity of an online four-target SSVEP decoding system outside standard laboratories by using a speed-controlled treadmill experiment that generated different degrees of head movements in moving humans. The online performance appeared to decrease as participants switched from standing to walking. A significant drop (p < 0.05) in accuracy and ITR was found during fast walking at 1.34 m/s (c.f. Figure 6). The reason might be attributed to the fact that the SSVEP amplitude was negatively correlated with the walking speed (c.f. Figures 3 and 5). The amplitude of SSVEP has been found to be largely modulated by visual spatial attention [28]. As a consequence, the loss of focus could reduce visual attention and thereby lead to the decreased SSVEP amplitude. This very likely happened in treadmill walking. Fast walking obviously initiated rapid head bouncing especially in fast walking (c.f. Figure 4). Such phenomenon was also confirmed by the participants of this study reported somehow more difficult in focusing on the flickering targets during walking, especially at the speed of 1.34 m/s. However, since the treadmill walking protocol adopted in this study did not randomize the walking speeds, we cannot completely rule out the possibility that the visual- or motor-fatigue could also contribute to the SSVEP deterioration. Currently, little is known about the interference of fatigue on SSVEP amplitude in an SSVEP experiment where multiple sessions are separated by inter-session breaks. A further study is needed to address the impact of fatigue on the BCI performance, which might likely happen in real-life applications.

Another interesting finding was that the increase of the gait cadence in our study (up to 1.34 m/s) accompanied larger head bobbing as well as moderate head swaying, resulting in head movements of 1–5 Hz (c.f. Figure 4). Such head movements inevitably swayed the headset and yielded a corresponding low-frequency power fluctuation spreading over the whole head (c.f. Figure 5), which was analogous to the evidence that gait-related artifacts significantly affected the EEG spectral power in the 1.5-8.5 Hz frequency range during walking (below 1.25 m/s) and running (1.9 m/s) [3]. Such 1–5 Hz mechanical artifacts produced by head movements during steady walking (tested up to 1.34 m/s in this study) should not affect the SSVEPs (9–12 Hz frequency range). In this regard, methods for movement artifact suppression [3] might not be able to improve the SSVEP detection performance in this study. However, the artifacts could contaminate other event-related potentials (ERPs) used in other types of BCIs. For example, P300 waveform is majorly composed of frontal theta (4-7 Hz) and posterior delta activations (<4 Hz) [29]. This gait-related low-frequency contamination could cause a deterioration of P300 quality during naturally walking [14, 17].

Lastly, one might concern the adopted treadmill-walking protocol prone to constrain the walking locomotion. This study instructed the subjects to hold the hand-grip while walking on the treadmill (c.f. Figure 1). It was not only to insure the subjects’ safety while proceeding with fast walking on the moving belt (1.34 m/s), but also to maintain their viewing distances to the display monitor while walking. Such hand-grip holding might result in dissimilar walking locomotion with less body and head movements compared to natural walking patterns in daily life. However, according to the measurements from the gyroscope (c.f. Figure 4), the treadmill-initiated walking locomotion evidently engaged pronounced head movements as compared to a standing posture. Unlike the previously laboratory-oriented BCI demonstrations with stationary subjects, this study was a valuable step towards the transition to real life applications.

Although the empirical results in this study demonstrated the applicability of using the consumer-level Emotiv headset to decode SSVEPs in moving humans, the obtained satisfactory performance might not assure that the headset could be perfectly suitable for all BCI applications. First, given the sparse 14-channel montage, the unavailability of midline and centro-parietal regions might hinder the measurements of good-quality P300 waveforms [14]. Second, the EEG quality measured by the designed saline electrodes highly depends on the moisture level of saline liquid, leading to progressive quality deterioration in a long-term use. Third, unlike the non-prep, dry electrodes [4], the pads of the saline electrodes might meet difficulties in having good contacts to hair-covered sites on the scalp. However, despite these potential drawbacks and inferior signal quality compared to medical-/laboratory-grade devices [30], the advantage of ease-to-use still could facilitate the BCI applications for non-critical applications in entertainment and game control.

Future efforts can be devoted to improve the online performance for freely moving humans. Wei et al.[31] recently proposed a new SSVEP detection method, namely differential canonical correlation analysis (dCCA), to improve the detectability of high-frequency SSVEPs as compared to the standard CCA. Thus, a natural next step is to employ the dCCA to improve the effectiveness of the SSVEP decoding. Another direction is to incorporate the consumer-level EEG headset with a head-mounted display device (presenting visual stimuli) to establish ubiquitous mobile BCI systems in ecologically valid environments.

Conclusions

This study implemented the first online SSVEP decoding system in walking humans using a low-cost, easy-to-use commercial EEG headset, providing principles towards translating laboratory-oriented BCI demonstrations to practical BCI applications in daily life. The empirical results showed that the Emotiv EEG headset enabled SSVEP decoding with ITRs above 12 bits/min during slow walking (below 0.89 m/s). The ITR reported in the present study is comparable to the previous studies with stationary subjects. For example, a four-class SSVEP-based BCI speller using a commercial EEG system obtained an average ITR of 13 bits/min across a total of 106 subjects [32]. This study evidently demonstrated that SSVEP-based BCI systems are deployable to users in natural head/body positions and movements rather than in severely restrict movements within highly controlled laboratory environments.

Abbreviations

BCI:

Brain-computer interface

CCA:

Canonical correlation analysis

dCCA:

Differential canonical correlation analysis

EEG:

Electroencephalography

ERP:

Event-related potential

FIR:

Finite impulse response

ITR:

Information transfer rate

MPH:

Miles per hour

PSD:

Power spectral density

SNR:

Signal-to-noise ratio

SSVEP:

Steady-state visual-evoked potential

STFT:

Short-time Fourier transform

m/s:

Meters per second

UCSD:

University of California San Diego.

References

  1. Gramann K, Gwin JT, Ferris DP, Oie K, Jung TP, Lin CT, Liao LD, Makeig S: Cognition in action: imaging brain/body dynamics in mobile humans. Rev Neuroscience. 2011, 22: 593-608.

    Article  Google Scholar 

  2. McDowell K, Lin CT, Oie KS, Jung TP, Gordon S, Whitaker KW, Li SY, Lu SW, Hairston WD: Real-World neuroimaging technologies. Access, IEEE. 2013, 1: 131-149.

    Article  Google Scholar 

  3. Gwin JT, Gramann K, Makeig S, Ferris DP: Removal of movement artifact from high-density EEG recorded during walking and running. J Neurophysiol. 2010, 103: 3526-3534.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Chi YM, Wang YT, Wang Y, Maier C, Jung TP, Cauwenberghs G: Dry and noncontact EEG sensors for mobile brain-computer interfaces. IEEE T Neur Sys Reh. 2012, 20: 228-235.

    Article  Google Scholar 

  5. Wang YT, Wang Y, Jung TP: A cell-phone-based brain-computer interface for communication in daily life. J Neural Eng. 2011, 8: 025018-025015pp

    Article  PubMed  Google Scholar 

  6. Wang Y, Wang YT, Jung TP: Visual stimulus design for high-rate SSVEP BCI. Electron Lett. 2010, 46: 1057-1058.

    Article  Google Scholar 

  7. Crowley K, Sliney A, Pitt I, Murphy D: Evaluating a Brain-Computer Interface to Categorise Human Emotional Response. Advanced Learning Technologies (ICALT), 2010 IEEE 10th International Conference on; 5–7 July 2010. 2010, 276-278.

    Google Scholar 

  8. Campbell A, Choudhury T, Hu S, Lu H, Mukerjee MK, Rabbi M, Raizada RDS: NeuroPhone: Brain-Mobile Phone Interface using a Wireless EEG Headset. Book NeuroPhone: Brain-Mobile Phone Interface using a Wireless EEG Headset. 2010, City: ACM, 3-8.

    Google Scholar 

  9. Bobrov P, Frolov A, Cantor C, Fedulova I, Bakhnyan M, Zhavoronkov A: Brain-computer interface based on generation of visual images. PloS One. 2011, 6: e20674-

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  10. Petersen MK, Stahlhut C, Stopczynski A, Larsen JE, Hansen LK: Smartphones Get Emotional: Mind Reading Images and Reconstructing the Neural Sources. Book Smartphones Get Emotional: Mind Reading Images and Reconstructing the Neural Sources. 2011, City: Springer-Verlag, 578-587.

    Google Scholar 

  11. Yue L, Xiao J, Teng C, Feng W, Peng Un M, Pui-In M, Mang IV: Implementation of SSVEP based BCI with Emotiv EPOC. Virtual Environments Human-Computer Interfaces and Measurement Systems (VECIMS), 2012 IEEE International Conference on; 2–4 July 2012. 2012, Tianjin, China: IEEE, 34-37.

    Google Scholar 

  12. Choi B, Jo S: A low-cost EEG system-based hybrid brain-computer interface for humanoid robot navigation and recognition. PloS One. 2013, 8: e74583-

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  13. Guneysu A, Akin HL: An SSVEP based BCI to Control a Humanoid Robot by using Portable EEG Device. Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE; 3–7 July 2013. 2013, Osaka, Japan: IEEE, 6905-6908.

    Chapter  Google Scholar 

  14. Debener S, Minow F, Emkes R, Gandras K, de Vos M: How about taking a low-cost, small, and wireless EEG for a walk?. Psychophysiology. 2012, 49: 1617-1621.

    Article  PubMed  Google Scholar 

  15. Lin YP, Wang Y, Wei CS, Jung TP: A Mobile Brain-Computer Interface for Freely Moving Humans. Human-Computer Interaction, Part V, HCII 2013, LNCS 8008. Edited by: Kurosu M. 2013, Berlin Heidelberg: Springer, 448-453.

    Google Scholar 

  16. Lotte F, Fujisawa J, Touyama H, Ito R, Hirose M, L\ A, \#233, cuyer: Towards Ambulatory Brain-Computer Interfaces: a Pilot Study with P300 Signals. Book Towards Ambulatory Brain-Computer Interfaces: a Pilot Study with P300 Signals. 2009, City: ACM, 336-339.

    Google Scholar 

  17. De Vos M, Gandras K, Debener S: Towards a truly mobile auditory brain-computer interface: exploring the P300 to take away. Int J Psychophysiol. 2014, 91 (1): 46-53.

    Article  PubMed  Google Scholar 

  18. Lin YP, Wang Y, Jung TP: A Mobile SSVEP-based Brain-computer Interface for Freely Moving Humans: The Robustness of Canonical Correlation Analysis to Motion Artifacts. the 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC’13); July 3–7. 2013, Osaka, Japan: IEEE, 1350-1353.

    Google Scholar 

  19. Bin GY, Gao XR, Yan Z, Hong B, Gao SK: An online multi-channel SSVEP-based brain-computer interface using a canonical correlation analysis method. J Neural Eng. 2009, 6: 046002-046006pp

    Article  PubMed  Google Scholar 

  20. Muller-Putz GR, Pfurtscheller G: Control of an electrical prosthesis with an SSVEP-based BCI. IEEE T Bio-Med Eng. 2008, 55: 361-364.

    Article  Google Scholar 

  21. Wang Y, Wang R, Gao X, Hong B, Gao S: A practical VEP-based brain-computer interface. IEEE T Neur Sys Reh. 2006, 14: 234-240.

    Article  CAS  Google Scholar 

  22. Renard Y, Lotte F, Gibert G, Congedo M, Maby E, Delannoy V, Bertrand O: L A, #233, cuyer: Openvibe: an open-source software platform to design, test, and use brain–computer interfaces in real and virtual environments. Presence: Teleoper Virtual Environ. 2010, 19: 35-53.

    Article  Google Scholar 

  23. Delorme A, Mullen T, Kothe C, Acar ZA, Bigdely-Shamlo N, Vankov A, Makeig S: EEGLAB, SIFT, NFT, BCILAB, and ERICA: new tools for advanced EEG processing. Intell Neuroscience. 2011, 2011: 10-10.

    Google Scholar 

  24. Lin Z, Zhang C, Wu W, Gao X: Frequency recognition based on canonical correlation analysis for SSVEP-based BCIs. IEEE T Bio-Med Eng. 2006, 53: 2610-2614.

    Article  Google Scholar 

  25. Jin J, Allison BZ, Sellers EW, Brunner C, Horki P, Wang XY, Neuper C: An adaptive P300-based control system. J Neural Eng. 2011, 8: 036006-

    Article  PubMed  PubMed Central  Google Scholar 

  26. Wolpaw JR, Birbaumer N, McFarland DJ, Pfurtscheller G, Vaughan TM: Brain-computer interfaces for communication and control. Clin Neurophysiol. 2002, 113: 767-791.

    Article  PubMed  Google Scholar 

  27. Delorme A, Makeig S: EEGLAB: an open source toolbox for analysis of single-trial EEG dynamics including independent component analysis. J Neurosci Meth. 2004, 134: 9-21.

    Article  Google Scholar 

  28. Morgan ST, Hansen JC, Hillyard SA: Selective attention to stimulus location modulates the steady-state visual evoked potential. Proc Natl Acad Sci U S A. 1996, 93: 4770-4774.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  29. Jones KA, Porjesz B, Chorlian D, Rangaswamy M, Kamarajan C, Padmanabhapillai A, Stimus A, Begleiter H: S-transform time-frequency analysis of P300 reveals deficits in individuals diagnosed with alcoholism. Clin Neurophysiol. 2006, 117: 2128-2143.

    Article  PubMed  Google Scholar 

  30. Duvinage M, Castermans T, Petieau M, Hoellinger T, Cheron G, Dutoit T: Performance of the Emotiv Epoc headset for P300-based applications. Biomed Eng Online. 2013, 12: 56-

    Article  PubMed  PubMed Central  Google Scholar 

  31. Wei CS, Lin YP, Wang Y, Wang YT, Jung TP: Detection of steady-state visual-evoked potential using differential canonical correlation analysis. the 6th International IEEE/EMBS Conference on Neural Engineering; 6–8 Nov. 2013. 2013, San Diego, USA: IEEE, 57-60.

    Chapter  Google Scholar 

  32. Allison B, Luth T, Valbuena D, Teymourian A, Volosyak I, Graser A: BCI demographics: how many (and What Kinds of) people can use an SSVEP BCI?. Neural Syst Rehabil Eng, IEEE Transactions. 2010, 18: 107-116.

    Article  Google Scholar 

Download references

Acknowledgments

This work was supported by Office of Naval Research (N00014-08-1215), Army Research Office (un-der contract number W911NF-09-1-0510), Army Research Laboratory (under Cooperative Agreement Number W911NF-10-2-0022), and DARPA (USDI D11PC20183). The authors acknowledge Christian Kothe for online programming assistance.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Tzyy-Ping Jung.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

YPL undertook the data collection, performed the data analysis, interpreted the results, and drafted the manuscript. YW conceived the experiments and helped to analyze data, interpret the results, and revised the manuscript. TPJ helped to interpret the results and revise the manuscript. All authors read and approved the final manuscript.

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver ( https://creativecommons.org/publicdomain/zero/1.0/ ) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Lin, YP., Wang, Y. & Jung, TP. Assessing the feasibility of online SSVEP decoding in human walking using a consumer EEG headset. J NeuroEngineering Rehabil 11, 119 (2014). https://doi.org/10.1186/1743-0003-11-119

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1743-0003-11-119

Keywords