Skip to main content

Comprehensive neuromechanical assessment in stroke patients: reliability and responsiveness of a protocol to measure neural and non-neural wrist properties

Abstract

Background

Understanding movement disorder after stroke and providing targeted treatment for post stroke patients requires valid and reliable identification of biomechanical (passive) and neural (active and reflexive) contributors. Aim of this study was to assess test-retest reliability of passive, active and reflexive parameters and to determine clinical responsiveness in a cohort of stroke patients with upper extremity impairments and healthy volunteers.

Methods

Thirty-two community-residing chronic stroke patients with an impairment of an upper limb and fourteen healthy volunteers were assessed with a comprehensive neuromechanical assessment protocol consisting of active and passive tasks and different stretch reflex-eliciting measuring velocities, using a haptic manipulator and surface electromyography of wrist flexor and extensor muscles (Netherlands Trial Registry number NTR1424).

Intraclass correlation coefficients (ICC) and Standard Error of Measurement were calculated to establish relative and absolute test-retest reliability of passive, active and reflexive parameters. Clinical responsiveness was tested with Kruskal Wallis test for differences between groups.

Results

ICC of passive parameters were fair to excellent (0.45 to 0.91). ICC of active parameters were excellent (0.88-0.99). ICC of reflexive parameters were fair to good (0.50-0.74). Only the reflexive loop time of the extensor muscles performed poor (ICC 0.18). Significant differences between chronic stroke patients and healthy volunteers were found in ten out of fourteen parameters.

Conclusions

Passive, active and reflexive parameters can be assessed with high reliability in post-stroke patients. Parameters were responsive to clinical status. The next step is longitudinal measurement of passive, active and reflexive parameters to establish their predictive value for functional outcome after stroke.

Background

Upper extremity movement disorder is a major contributor to impaired activity and participation levels in post-stroke patients [1,2]. In the acute phase after stroke, paresis is the dominant factor of impairment [3,4]. However, in the chronic phase, the complex interaction between inappropriate neural activation of muscles and secondary biomechanical changes in contractile muscle tissue and passive viscoelastic connective tissue becomes more prominent [3-6]. The dynamical interactions between neural capacity and contractile and connective tissues during daily functioning in patients are poorly understood.

Unraveling movement disorder after stroke into non-neural (passive) and neural (active and reflexive) contributors and assess their separate influence over time, is essential to understand functional recovery after stroke and to aim therapy at the most dominant contributing factor at the most appropriate moment in time [7-10]. Physical examination is currently the most utilized clinical tool for assessment of paresis, inappropriate muscle activity and secondary biomechanical changes [11].

Biomechanical and electrophysiological techniques support standardization of input signals and uniform registration of output signals. A comprehensive neuromechanical assessment should be able to discriminate between non-neural and neural contributors to movement disorder [7-10,12]. Non-neural contributors, i.e. passive tissue properties, should be measured by passive movement at low velocity to minimize background muscle activation [13]. Neural contributors should be measured during active tasks to study voluntary muscle properties and during multiple measurement velocities to study the role of stretch reflexes [3,4,7-10,12]. System Identification and Parameter Estimation techniques assist in separation of neural and non-neural contributors independently of task and condition [14].

Earlier work on measurement of joint neuromechanics [15-19] provided a comprehensive assessment protocol including passive, active and reflexive tests to measure non-neural and neural contributors to movement disorder after stroke [20]. To ensure standardized input signals and registration of output signals, a haptic wrist manipulator [21,22] was combined with surface-EMG measurements.

Clinical implementation of this newly developed protocol required validation. The aim of this study was to assess test-retest reliability and to determine clinical responsiveness in a cohort of stroke patients with upper extremity impairments compared to a cohort of healthy volunteers.

Methods

Participants

We identified patients who survived a first ischemic stroke between 1999–2009, and were between 18–80 years at time of stroke, at the outpatient clinics of the Department of Rehabilitation in LUMC and Rijnlands Rehabilitation Center. Inclusion criteria were: a perceived remaining impairment of arm-hand function by the participant, being able to travel to the research laboratory, and being able to sit on a chair and follow instructions for one hour. To establish the perceived impairment of arm-hand function, respondents were asked if they still perceived any impairment of the arm and/or hand. Possible answers were: no impairment, moderate impairment or severe impairment. Patients with moderate to severe perceived impairment were invited for measurements. Exclusion criteria were: limitations of arm-hand function prior to stroke, a history of other neurologic impairments besides stroke. Participants were measured on two occasions within a month, assuming that their clinical status would remain stable. A volunteer sample of healthy volunteers served as a reference group. The study was approved by the Medical Ethical Committee of the LUMC. All participants were compensated for travel expenses.

Protocol

Measurements were carried out in a laboratory setting at the LUMC. Before the test protocol started, the modified Ashworth Scale (mAS) was measured. Participants were extensively instructed and were given ample opportunity to practice. The protocol consisted of nine tests, with a total duration of approximately 45 minutes.

Measurement set-up

The measurement set-up [20] consisted of a haptic manipulator (“Wristalyzer”, Moog FCS, the Netherlands) and a surface EMG-system (“Bagnoli”, Delsys Inc., USA). The manipulator delivered precise torque or position perturbations through a vertically positioned servo-motor (Parker SMH100 series), connected to a handle (Meester techniek, the Netherlands). The hand of the participant was fixed to the handle, which had an ellipsoidal shape to prevent finger flexion (Figure 1). The arm was stabilized in an arm rest. The motor axis was aligned with the rotation axis of the wrist joint. Movement of the motor was therefore directly coupled to flexion/extension of the wrist.

Figure 1
figure 1

Illustration of Wristalyzer handle and arm-rest. For a better view of the hand position, the hand straps are not represented in this illustration.

Tasks, conditions and outcome parameters

An overview and more elaborate description of applied Passive, Active and Reflexive tests and their outcome parameters is shown in Table 1. The tests were performed in a fixed order, starting with Passive tests, followed by Active and Reflexive tests respectively [20]. Participants were provided with visual feedback on torque, angle or EMG-level, depending on the test and task instruction.

Table 1 Description of passive, active and reflexive parameters

Passive tests were performed at low velocity to avoid stretch reflexes, and included a task instruction to “do nothing”. First, a force controlled movement was applied in flexion and extension direction, to establish passive range of motion passive (PROM). Then a position controlled movement was applied, also in both flexion and extension direction. The following outcome parameters were extracted: Stiffness in rest (Pk) and Rest angle (PRA) (Table 1).

Active tests comprised task instructions to “move/push/resist”, i.e. exert a voluntary torque or complete a voluntary movement. This part commenced with an active, maximal movement from flexion to extension and back to establish active range of motion active (AROM). Then, the position of the handle was fixed at the PRA and participants were asked to complete and repeat a maximal voluntary isometric contraction in both flexion and extension direction to establish active maximal voluntary contraction (AMVC). Subsequently, participants were asked to match a gradually inclining torque level. This was also performed in both flexion and extension direction and repeated once. The following outcome parameter was extracted: Control over joint torque (ACJT) (Table 1).

Reflexive tests were performed at velocities above the assumed stretch reflex threshold, and had either passive or active task instructions. A high velocity, position controlled movement through the PROM was applied once, in both flexion and extension direction (passive task instruction) to calculate Threshold angle (Rta). Short ramp and hold position perturbations were applied 9 times in each direction at random time intervals (active task instruction) to extract Reflexive loop time (Rlt). A multisine force perturbation was applied for 20 seconds (active task instruction) and repeated three times: once in the same environment and twice in a damped environment. The following outcome parameters were extracted: Reflexive contributions to joint resistance (Rkv) and Reflex modulation due to environmental changes (Rm_env) (Table 1).

Statistical analysis

Data was retrieved and processed with Matlab 2007b (Mathworks, USA) [20]. Calculations and statistics were performed with SPSS Statistics 20 (IBM, USA). Sample size was calculated on the outcome parameter with the expected largest variability: Rkv. In an earlier study, a standard deviation of 0.17 Nms/rad was found [17], with a mean difference between patients and controls of 0.12 Nms/rad. Based upon α = 0.05 and with a target power of 80%, a sample size of minimally 10 participants was estimated to be required to detect an existing difference between measurements of 0.12 Nms/rad with sufficient power.

Intraclass correlation coefficients (ICC) were calculated using the two-way mixed model for absolute agreement. Values above 0.75 represent excellent reliability, values between 0.4 and 0.75 represent fair to good reliability and values below 0.4 represent poor reliability [23]. As ICC is a relative measure dependent on variance within a group [24], Bland Altman plots were used to illustrate variability and Standard Error of Measurement (SEM) values (Equation 1) and Smallest Detectable Difference (SDD) (Equation 2) were calculated to further substantiate ICC.

$$ SEM = SD*\surd \left(1-ICC\right) $$
(1)
$$ SDD = 1.96\ *\ \surd 2*SEM $$
(2)

Normality of distribution was assessed by visual inspection of histograms and equality of variance was tested with Levene’s test. Median, minimum and maximum were calculated per parameter. Levene’s test showed wider variances in the group of chronic stroke patients compared to healthy volunteers, therefore chronic stroke patients were split in two groups according to mAS (mAS = 0 and mAS ≥ 1), a clinimetric observation. This allowed for a more specific description of phenotypes and more equally dispersed values within each group, which was illustrated with box plots. An exploratory comparison was made between parameters of the group of healthy volunteers, the group of chronic patients with mAS = 0 and the group of chronic patients with mAS ≥ 1, using the non-parametric Kruskal-Wallis test. The Kruskal-Wallis test was performed on the average outcome of the two visits per parameter, after testing for systematical differences between the two visits. This comparison was further substantiated by pairwise testing between each of the groups with the non-parametric Wilcoxon Rank Sum test.

Results

Descriptive data

We identified and invited 102 post stroke patients. Response rate was 64% (n = 65). Of the responders, 17 patients declined to participate and 16 patients had either no current impairment of arm-hand function or had severely impaired mobility preventing them from travelling to the clinic. Therefore, 32 patients were included in the study. Fourteen healthy volunteers served as a reference group. All healthy volunteers completed the two visits and 28 out of 32 patients completed all visits (87.5%). Reasons for dropping out were: unable to schedule the second visit (n = 2), visit was too tiresome (n = 1), patient was treated with botulinum toxin in period between first and scheduled second visit (n = 1). In chronic stroke patients, the affected hand was dominant in 14 patients (right hand n = 13, left hand n = 1) and non-dominant in 18 patients (right hand n = 2, left hand n = 16). Average age at stroke was 55.2 years. Average time after stroke was 40 months. Descriptive data are presented in Table 2.

Table 2 Descriptive data of the study population

Reliability

Test-retest reliability for Passive parameters was excellent for PROM and Pk (ICC 0.81 and 0.91 respectively) and fair to good for PRA (ICC 0.45). For Active parameters, test-retest reliability was excellent (ICC 0.88-0.99). For Reflexive parameters, the ICC’s of Rta (flexor and extensor), Rlt (flexor), Rkv and Rm_env were fair to good (ICC 0.50-0.74). ICC for Rlt (extensor) was poor (ICC 0.18). ICC’s are summarized in Table 3. Bland Altman plots are shown in Figure 2, depicting the mean of the two measurements (x-axis) compared to the difference between two measurements (y-axis). The values are scattered around the mean difference (solid line), illustrating the absence of a systematic difference or learning effect between the two measurements. In parameters with a lower ICC, the 95% confidence interval of the difference between the measurements is wider, illustrating a larger measurement error. SEM values (Table 3) provide an indication of the dispersion of the measurement errors and SDD are given for future reference (Table 3).

Table 3 Median, minimum and maximum, intraclass correlation coefficients (ICC), standard errors of measurement (SEM) and smallest detectable difference (SDD) for passive, active and reflexive parameters for all participants
Figure 2
figure 2

Bland altman plots for passive, active and reflexive parameters. PROM: Range of motion passive, Pk: Stiffness in rest, PRA: Rest angle. AROM: Range of motion active, AMVC: Maximal voluntary contraction, ACJT: Control over joint torque. Rta: Threshold angle, Rlt: Reflexive loop time, Rkv: Reflexive contributions to joint resistance, Rm_env: Reflex modulation due to environmental changes. Open circle: Healthy participant. Gray circle: Post-stroke patient. Solid line: mean of the difference between first en second visits. Dotted line: upper and lower limit of 95% confidence interval for difference between first and second visits.

Responsiveness to clinical status

An overview of outcomes per group is summarized in Table 4. Dropout rates were similar in both chronic patient groups (mAS = 0: n = 2 and mAS ≥ 1: n = 2). Differences between healthy volunteers, chronic stroke patients with mAS = 0 and chronic stroke patients with mAS ≥ 1 are illustrated by box plots in Figure 3. Corresponding quartiles are given in Additional file 1. In 10 out of 14 parameters, these differences were statistically significant, based on the exploratory Kruskal Wallis test (Table 4). When tested pairwise with the Wilcoxon Rank Sum test (Additional file 2), there were significant differences between healthy volunteers and the mAS = 0 group in the Passive parameters Pk and PRA; the Active parameters AROM, AMVC (flexor and extensor) and ACJT (flexor and extensor); and the Reflexive parameters Rlt (extensor) and Rkv. Differences between healthy volunteers and the mAS ≥ 1 group showed significance in PROM, all Active parameters, Rta (extensor) and Rkv. When comparing the mAS = 0 group and mAS≥1 group, there were significant differences in PROM, Pk and all Active parameters, but no significant differences in Reflexive parameters.

Table 4 Median, minimum and maximum for healthy volunteers, chronic patients with modified Ashworth score (mAS) = 0 and chronic patients with mAS ≥ 1, and p-value of the Kruskal Wallis test for differences between groups for passive, active and reflexive parameters
Figure 3
figure 3

Box plots for Passive, Active and Reflexive parameters. Groups divided in healthy volunteers, chronic stroke patients with mAS = 0 and chronic patients with mAS ≥1. PROM: Range of motion passive, Pk: Stiffness in rest, PRA: Rest angle. AROM: Range of motion active, AMVC: Maximal voluntary contraction, ACJT: Control over joint torque. Rta: Threshold angle, Rlt: Reflexive loop time, Rkv: Reflexive contributions to joint resistance, Rm_env: Reflex modulation due to environmental changes. White bars: flexor (and Rkv in the lower right panel). Striped bars: extensor (and Rm_env in the lower right panel).

Discussion

Using a dedicated, comprehensive neuromechanical assessment protocol, Passive and Active parameters could be assessed with excellent reliability in a cohort of stroke patients with upper extremity impairments and healthy volunteers. Repetitive assessment of the Passive parameter PRA and Reflexive parameters had fair to good reliability, except for poor reliability of Rlt (extensor). Parameters were responsive to clinical status, i.e. results demonstrated differences between healthy volunteers and chronic stroke patients.

The use of a haptic robot in combination with surface EMG provides standardized application of input signals and registration of output signals. Participants could comfortably tolerate the position in the measurement set-up and the length of the protocol. Previous publication of measurement set-up, protocol and data processing [20] and current assessment of test-retest reliability and clinical responsiveness add to the clinical validity of our method, which is advantageous for prospective implementation of this method in clinical practice.

Relative reliability expressed by ICC’s is both determined by heterogeneity of the study group and the variance on the repeated measurements. In homogenous study groups, relative reliability may drop. For future assessment of longitudinal changes, variability of neuromechanical outcome parameters is unknown and may be dependent on the time of measurement, i.e. low heterogeneity early after stroke when the paresis component prevails and large heterogeneity when secondary biomechanical changes become manifest. We therefore adopted the present cohort-approach to minimize a-priori assumptions on heterogeneity within groups.

Measures of absolute reliability can be used to calculate the SDD, i.e. the difference between measurements that can be attributed to real system changes. For example in AROM the SDD is 14 degrees, meaning that a change of 14 degrees or over can be attributed to a genuine change in patient characteristics in 95% of cases. SEM values for PROM, Pk, AROM, AMVC and ACJT were low. These are sensitive parameters for real system changes, indicative for both passive as well as active contributors to observed movement disorders. We therefore recommend these parameters for future assessment of longitudinal neuromechanical changes.

In all Passive, Active and Reflexive parameters except Pk, PRA and Rlt (extensor), differences between healthy volunteers and patients in the mAS ≥ 1 group were more pronounced than differences between healthy volunteers and patients in the mAS = 0 group. However, parameters did not always increase or decrease proportionally between groups, which illustrates the complex and non-linear nature of movement disorder after stroke. In the mAS = 0 group, the paresis component probably plays an important role, while the ability for voluntary motor control is more preserved than in the mAS ≥ 1 group (as can be seen from AMVC and ACJT), leading to a lower stiffness (Pk) in passive structures (i.e. muscle, tendon, ligament). Test-retest results showed good reproducibility, however, the remarkable inter-individual variation in passive and active parameters in the group of chronic patients may represent the different phenotypes in post stroke motor control.

Strengths and limitations

The perturbations in our protocol may not have been enough to trigger the stretch reflex threshold of the extensor muscles, which are more difficult to trigger [25]. This could have contributed to a lower repeatability in Reflexive parameters Rlt (extensor) and Rta (extensor), and a larger SDD than expected for Rkv. Other contributing factors may be found in a low signal to noise ratio, i.e. absence of inappropriate muscle activity in healthy volunteers and in chronic stroke patients with mAS = 0. Furthermore, variability in Reflexive parameters is known to be present in both healthy volunteers and chronic patients [26-29], even in optimal circumstances. Stretch reflex behavior is more variable than passive tissue properties and voluntary muscle properties [29-31]. Apart from day-to-day variability in stretch reflex behavior [27,29], there is also variability due to level of arousal, audiovisual stimuli and other environmental factors [32,33], as well as conscious down- or up-regulation [34]. These methodological considerations, combined with the unequal variances in subgroups, might account for heteroscedasticity, especially in Rta flexor and Rlt extensor. SEM and SDD for these values should be interpreted bearing in mind that SDD and SEM might be smaller towards the minimum of the parameter and might be larger towards the maximum of both parameters.

Although sufficient for the aim of this study, group sizes were small. Current subdivision of clinical phenotypes according to mAS is a fairly rough approximation of clinical status and more participants may be needed for a more elaborate post-hoc analysis. The adopted cohort approach is an estimation of group heterogeneity. The neuromechanical assessment protocol aimed to identify passive, active and reflexive contributors to movement disorder by differences in task and measurement conditions. For example: the protocol was designed to minimize the effects of active and reflexive (neural) contributors during passive (non-neural) tasks and vice versa. However, this might not yet give a completely true reflection of neuromechanical behavior, as system behavior under active task conditions involves a combination of both neural and non-neural contributors. The same goes for passive conditions, where neural components may be present through increased baseline activation [13]. Further development of System Identification and Parameter Estimation techniques might help to zoom in even closer on the specific contributors to neuromechanical behavior.

Implications for future work

One of the objectives of the EXPLICIT-stroke project [35] will be to combine the neuromechanical approach with extensive clinimetric data. Simultaneously, in this project, longitudinal measurements will be used to provide information on the changes in paresis, the development of secondary biomechanical changes and the increase of inappropriate muscle activity over time. This should provide the necessary data to enhance description of clinical phenotypes by clustering of neuromechanical parameters, and, moreover, to predict functional outcome.

Conclusions

Passive, Active and Reflexive parameters, representing passive tissue properties, voluntary muscle function and stretch reflex behavior respectively, can be measured in a reliable way. The comprehensive neuromechanical assessment protocol is responsive to clinical status and fulfills the requirements to separately assess non-neural and neural contributors to movement disorder around the wrist after stroke, using biomechanical, electromyographical and system identification techniques [7-10,12]. Therefore, this protocol gives momentum to future work on connecting pathophysiology to functional outcome, which will enable clinicians to substantiate their treatment.

Abbreviations

LUMC:

Leiden University Medical Center

mAS:

Modified Ashworth Scale

ICC:

Intraclass correlation coefficient

SEM:

Standard error of measurement

SDD:

Smallest detectable difference

EMG:

Electromyography

References

  1. Dobkin BH. Clinical practice. Rehabilitation after stroke. N Engl J Med. 2005;352:1677–84.

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  2. Kwakkel G, Kollen BJ, van der Grond J, Prevo AJ. Probability of regaining dexterity in the flaccid upper limb: impact of severity of paresis and time since onset in acute stroke. Stroke. 2003;34:2181–6.

    Article  PubMed  Google Scholar 

  3. Gracies JM. Pathophysiology of spastic paresis. II: Emergence of muscle overactivity. Muscle Nerve. 2005;31:552–71.

    Article  PubMed  Google Scholar 

  4. Gracies JM. Pathophysiology of spastic paresis. I: Paresis and soft tissue changes. Muscle Nerve. 2005;31:535–51.

    Article  PubMed  Google Scholar 

  5. Dietz V, Sinkjaer T. Spastic movement disorder: impaired reflex function and altered muscle mechanics. Lancet Neurol. 2007;6:725–33.

    Article  PubMed  Google Scholar 

  6. Sommerfeld DK, Gripenstedt U, Welmer AK. Spasticity after stroke: an overview of prevalence, test instruments, and treatments. Am J Phys Med Rehabil. 2012;91:814–20.

    Article  PubMed  Google Scholar 

  7. Burridge JH, Wood DE, Hermens HJ, Voerman GE, Johnson GR, Van Wijck FM, et al. Theoretical and methodological considerations in the measurement of spasticity. Disabil Rehabil. 2005;27:69–80.

    Article  CAS  PubMed  Google Scholar 

  8. Pandyan AD, Gregoric M, Barnes MP, Wood D, Van Wijck FM, Burridge J, et al. Spasticity: clinical perceptions, neurological realities and meaningful measurement. Disabil Rehabil. 2005;27:2–6.

    Article  CAS  PubMed  Google Scholar 

  9. Platz T, Eickhof C, Nuyens G, Vuadens P. Clinical scales for the assessment of spasticity, associated phenomena, and function: a systematic review of the literature. Disabil Rehabil. 2005;27:7–18.

    Article  CAS  PubMed  Google Scholar 

  10. Wood DE, Burridge JH, Van Wijck FM, McFadden C, Hitchcock RA, Pandyan AD, et al. Biomechanical approaches applied to the lower and upper limb for the measurement of spasticity: a systematic review of the literature. Disabil Rehabil. 2005;27:19–32.

    Article  CAS  PubMed  Google Scholar 

  11. Gracies JM, Bayle N, Vinti M, Alkandari S, Vu P, Loche CM, et al. Five-step clinical assessment in spastic paresis. Eur J Phys Rehabil Med. 2010;46:411–21.

    PubMed  Google Scholar 

  12. Van der Krogt JM, Meskers CG, De Groot JH, Klomp A, Arendzen JH. The gap between clinical gaze and systematic assessment of movement disorder after stroke. J Neuroeng Rehabil. 2012;9:61.

    Article  PubMed Central  PubMed  Google Scholar 

  13. Burne JA, Carleton VL, O’Dwyer NJ. The spasticity paradox: movement disorder or disorder of resting limbs? J Neurol Neurosurg Psychiatry. 2005;76:47–54.

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  14. Perreault EJ, Crago PE, Kirsch RF. Estimation of intrinsic and reflex contributions to muscle dynamics: a modeling study. IEEE Trans Biomed Eng. 2000;47:1413–21.

    Article  CAS  PubMed  Google Scholar 

  15. De Vlugt E, Schouten AC, van der Helm FC. Adaptation of reflexive feedback during arm posture to different environments. Biol Cybern. 2002;87:10–26.

    Article  PubMed  Google Scholar 

  16. De Vlugt E, De Groot JH, Schenkeveld KE, Arendzen JH, van der Helm FCT, Meskers CGM. The relation between neuromechanical parameters and Ashworth score in stroke patients. J Neuroeng Rehabil. 2010;7:35.

    Article  PubMed Central  PubMed  Google Scholar 

  17. Meskers CG, Schouten AC, De Groot JH, De Vlugt E, Van Hilten BJ, van der Helm FC, et al. Muscle weakness and lack of reflex gain adaptation predominate during post-stroke posture control of the wrist. J Neuroeng Rehabil. 2009;6:29.

    Article  PubMed Central  PubMed  Google Scholar 

  18. Schuurmans J, De Vlugt E, Schouten AC, Meskers CG, De Groot JH, van der Helm FC. The monosynaptic Ia afferent pathway can largely explain the stretch duration effect of the long latency M2 response. Exp Brain Res. 2009;193:491–500.

    Article  PubMed  Google Scholar 

  19. Van der Helm FC, Schouten AC, De Vlugt E, Brouwn GG. Identification of intrinsic and reflexive components of human arm dynamics during postural control. J Neurosci Methods. 2002;119:1–14.

    Article  PubMed  Google Scholar 

  20. Klomp A, Van der Krogt JM, Meskers CGM, De Groot JH, de Vlugt E, van der Helm FCT, et al. Design of a concise and comprehensive protocol for post stroke neuromechanical assessment. J Bioeng Biomed Sci. 2012;S1:008.

    Google Scholar 

  21. Grimaldi G, Lammertse P, Van Den Braber N, Meuleman J, Manto M. A New Myohaptic Device to Assess Wrist Function in the Lab and in the Clinic - The Wristalyzer. In: Ferre M, editor. EuroHaptics. Berlin Heidelberg: Springer-Verlag; 2008. p. 33–44.

    Google Scholar 

  22. Schouten AC, De Vlugt E, Van Hilten JJ, van der Helm FC. Design of a torque-controlled manipulator to analyse the admittance of the wrist joint. J Neurosci Methods. 2006;154:134–41.

    Article  PubMed  Google Scholar 

  23. Fleiss JL. Reliability of Measurement. In: Design and Analysis of Clinical Experiments. Wiley Interscience. 1999. p. 1–32.

  24. Hopkins WG. Measures of reliability in sports medicine and science. Sports Med. 2000;30:1–15.

    Article  CAS  PubMed  Google Scholar 

  25. Wargon I, Lamy JC, Baret M, Ghanim Z, Aymard C, Penicaud A, et al. The disynaptic group I inhibition between wrist flexor and extensor muscles revisited in humans. Exp Brain Res. 2006;168:203–17.

    Article  CAS  PubMed  Google Scholar 

  26. Blackburn JT, Mynark RG, Padua DA, Guskiewicz KM. Influences of experimental factors on spinal stretch reflex latency and amplitude in the human triceps surae. J Electromyogr Kinesiol. 2006;16:42–50.

    Article  PubMed  Google Scholar 

  27. Schmit BD, Dewald JP, Rymer WZ. Stretch reflex adaptation in elbow flexors during repeated passive movements in unilateral brain-injured patients. Arch Phys Med Rehabil. 2000;81:269–78.

    Article  CAS  PubMed  Google Scholar 

  28. Stam J, Van Crevel H. Measurement of tendon reflexes by surface electromyography in normal subjects. J Neurol. 1989;236:231–7.

    Article  CAS  PubMed  Google Scholar 

  29. Starsky AJ, Sangani SG, McGuire JR, Logan B, Schmit BD. Reliability of biomechanical spasticity measurements at the elbow of people poststroke. Arch Phys Med Rehabil. 2005;86:1648–54.

    Article  PubMed  Google Scholar 

  30. Lebiedowska MK, Fisk JR. Knee resistance during passive stretch in patients with hypertonia. J Neurosci Methods. 2009;179:323–30.

    Article  PubMed  Google Scholar 

  31. Mirbagheri MM, Barbeau H, Kearney RE. Intrinsic and reflex contributions to human ankle stiffness: variation with activation level and position. Exp Brain Res. 2000;135:423–36.

    Article  CAS  PubMed  Google Scholar 

  32. Stam J, Tan KM. Tendon reflex variability and method of stimulation. Electroencephalogr Clin Neurophysiol. 1987;67:463–7.

    Article  CAS  PubMed  Google Scholar 

  33. Turk R, Notley SV, Pickering RM, Simpson DM, Wright PA, Burridge JH. Reliability and Sensitivity of a Wrist Rig to Measure Motor Control and Spasticity in Poststroke Hemiplegia. Neurorehabil Neural Repair. 2008;22:684–96.

    Article  PubMed  Google Scholar 

  34. Wolf SL, Segal RL. Reducing human biceps brachii spinal stretch reflex magnitude. J Neurophysiol. 1996;75:1637–46.

    CAS  PubMed  Google Scholar 

  35. Kwakkel G, Meskers CG, Van Wegen EE, Lankhorst GJ, Geurts AC, Van Kuijk AA, et al. Impact of early applied upper limb stimulation: the EXPLICIT-stroke programme design. BMC Neurol. 2008;8:49.

    Article  PubMed Central  PubMed  Google Scholar 

Download references

Acknowledgements

EXPLICIT-stroke is funded by ZonMw (The Netherlands Organization for Health Research and Development) grant 890000001, Vereniging van Revalidatieartsen, Revalidatiefonds and Revalidatie Nederland. We would like to thank PH Goossens, MD PhD, at the Rijnlands Rehabilitation Centre, for her assistance in identifying participants for the study.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Hanneke van der Krogt.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

HK participated in the design of the study, carried out the measurements, performed the statistical analysis and drafted the manuscript. AK participated in the design of the study, carried out the measurements and helped to draft the manuscript. JG participated in the design of the study and its coordination, and helped to draft the manuscript. EV participated in the design of the study and commented on the manuscript. FH participated in the design of the study and commented on the manuscript. CM conceived of the study, and participated in its design and coordination and helped to draft the manuscript. JA conceived of the study, and participated in its design and coordination and helped to draft the manuscript. All authors read and approved the final manuscript.

Additional files

Additional file 1:

Quartiles of Passive, Active and Reflexive parameters for healthy volunteers, chronic patients with modified Ashworth score (mAS) = 0 and chronic patients with mAS ≥ 1. PROM: Range of motion passive, Pk: Stiffness in rest, PRA: Rest angle. AROM: Range of motion active, AMVC: Maximal voluntary contraction, ACJT: Control over joint torque. Rta: Threshold angle, Rlt: Reflexive loop time, Rkv: Reflexive contributions to joint resistance, Rm_env: Reflex modulation due to environmental changes.

Additional file 2:

Pairwise comparison with Wilcoxon Rank Sum test between healthy volunteers, chronic patients with mAS = 0 and chronic patients with mAS ≥ 1. #: significant difference between pair. The Kruskal Wallis test results are repeated from Table 4 for reference. PROM: Range of motion passive, Pk: Stiffness in rest, PRA: Rest angle. AROM: Range of motion active, AMVC: Maximal voluntary contraction, ACJT: Control over joint torque. Rta: Threshold angle, Rlt: Reflexive loop time, Rkv: Reflexive contributions to joint resistance, Rm_env: Reflex modulation due to environmental changes.

Rights and permissions

This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

van der Krogt, H., Klomp, A., de Groot, J.H. et al. Comprehensive neuromechanical assessment in stroke patients: reliability and responsiveness of a protocol to measure neural and non-neural wrist properties. J NeuroEngineering Rehabil 12, 28 (2015). https://doi.org/10.1186/s12984-015-0021-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12984-015-0021-9

Keywords