Resolution: standard / high Figure 1.
Control system architecture. The Cognitive Vision System (CVS) is integrated into a hierarchical control system
for the control of a dexterous prosthetic hand (emulated by the CyberHand prototype).
The user triggers the system and controls the orientation of the hand. A high-level
controller autonomously selects the grasp type and size that are appropriate for the
target object. A low-level controller embedded into the hand provides a stable interface
for preshaping and grasping.
Došen et al. Journal of NeuroEngineering and Rehabilitation 2010 7:42 doi:10.1186/1743-0003-7-42 |